Technical Parameter1. Range: (0-70) MPa
2. Medium: 25 # hydraulic oil
3. Minimum controllable adjustment: 10KPa
4. Interface size: 2 pieces/M20 × 1.5 Standard interface
Overview of Features
At present, piston pressure gauges, desktop pressure sources, electric pressure calibration tables, and other products are
commonly used in the metrological verification or calibration industry in China. These products are generally large in size and should not be carried on site for calibration. However, portable products from abroad use pliers. Due to the high pressure and complex operation, our company has recently developed a portable liquid pressure source that is easy to carry, simple in structure, and labor-saving in operation.
Points for attention
1.Instruments and meters with impurities must be cleaned before calibration.
2.When connecting instruments containing highly
corrosive media to this equipment, they must be cleaned with carbon tetrachloride solution.
3.When the threaded end face of the interface of the tested instrument is damaged, there may be leakage at the output port. Please configure the transition joint and sealing aluminum gasket produced by our company.
4. When not in use, please close each output port, tighten the oil injection port, and store in a clean and dry place.
5. Due to the presence of a large amount of impurities and oxides in the surface cavity of the inspected meter, it is recommended to purchase our company's liquid filter.
6. It is strictly prohibited to mix thisequipment with other equipment and use it beyond the rated pressure of this equipment.